NE8-15S
15 Litre Incubator, Fan Assist Circulation
DimensionsExternal: 515w x 440d x 550h mm
Internal: 320w x 180d x 240h mm
Packed:: 85x75x75cm
Capacity 15 litre chamber
Working capacity of the chamber can be smaller.
Temperature Range
ambient +5°C to 100°C
Sensitivity ±0.5°C @ 37°C / ±0.5°C @ 50°C
Sensitivity measurements based on temperature readings on shelf in centre of the chamber.
Uniformity±0.5°C @ 37°C
Uniformity measured in vertical axis in the chamber using Wired Shelves.
Door configuration
Single door Internal glass door and outer solid door with door lock and open door alarm
Shelves fitted 1
max. no.2Power consumption300WVoltage230V 50Hz 230V 60Hz and 115V 60Hz
Packed Weight27kg
SPECIFICATION ACCESSORIES MANUALS
Product Code NE8-15S
Description
Clifton Incubators feature intuitive easy to use controls with on screen guidance within a bold illuminated LCD display, offering precise time and temperature control from ambient +5°C to 100°C.
The Incubator can run at a set temperature immediately, delay start or commence to start in the future at a given date and time. Additionally with three available programs with up to 6 segments in each program which allows the creation of temperature time profiles of your choice.
All control surfaces feature smooth easy to wipe clean surfaces.
The digital microprocessor controls feature real time clock, diagnostics, digital LCD display to 0.1°C resolution, dual pt100 probes, including event log and data log. Every unit is supplied with a Calibration Certificate for validation.
Featuring a stainless steel interior with galvanised steel outer case powder painted. Particular attention is paid to providing a good tight door closure and hinge design to maintain a sealed chamber in use. Case is well insulated to reduce heat loss in use. Access port fitted as standard.
Controls Featuring bold illuminated LCD graphic display of measured temperatures in real-time, in increments of ±0.1°C, with fan speed status, set temperature, date and time.
Intuitive to use controls, utilising illuminated touch sensitive keypads with display ‘prompts’ assisting the setting of functions within the unit.
Option to set temperature in °C or °F.
The Incubator can be set to run at a set temperature immediately, delay start or commence to start in the future at a given time and date/year
With three available programs and up to 6 segments in each it allows the creation of temperature time profiles of your choice, including for example fan speed, exhaust air flap setting, fast achieve set temperature and a setting allowing programs to be cycled. Any of these programs they can run immediately, delay start or commence to start in the future at a given time and date]. And when a program is running the controls are locked preventing accidental interruption or change however stopping a program is straightforward.
Fan speed can be set from 0% no air convection within the chamber to 100% for maximum air convection.
Exhaust air flap can be set from 0% closed to 100% open and allows air from the chamber to vent.